Autoclave presoclave-iii 50l 220v 50/60hz

Autoclaves for sterilization “Presoclave” 220V 50/60Hz

Microprocessor temperature and cycle control and atmospheric purge.
Electronic control of time and temperature range from 115 °c to 134 °c. (0.62 to 2 bar). Conforms to apparatus pressure regulation directives.
Vertical model
capacity: 50 litres

PRINCIPAL OF STEAM STERILIZATION

STERILIZATION IS THE PROCESS OF THE DESTRUCTION AND ELIMINATION OF ALL FORMS OF MICROBIAL LIFE, INCLUDING BACTERIA AND ITS SUBFORMS.
THE LABORATORY AUTOCLAVE CONSISTS OF A VESSEL IN WHICH STEAM FLOWS AT A HIGH PRESSURE WITH A STERILIZATION PURPOSE. SATURATED STEAM WILL IMMEDIATELY CONDENSE AND THUS RELEASING ITS HIGH EVAPORATION HEAT, WHICH IS IN FACT THE DESTRUCTIVE AGENT. IN ORDER THAT STERILIZATION WOULD BE EFFECTIVE, STEAM MUST BE SATURATED AND, CONSEQUENTLY, ALL THE AIR INSIDE THE VESSEL MUST BE PURGED BEFORE INITIALIZING THE PROCESS.

CONTROL PANEL

1. Pressure gauge.
2. Illuminated alarm indicator.
3. Illuminated indicator door open.
4. Temperature display.
5. Time display.
6. Temperature set pushbutton (range from 115 °C to 134 °C).
7. Pushbutton increase value.
8. Push button START
9. Push Button STOP
10. Push button Reduce value.
11. Push button select time, (range from 3’ to 99’).
12. Push button open lid.
13. Devaporization pushbutton. 

BACK VIEW:

1. Safety valve.
2. Mains power.
3. Safety thermostat.
4. Steam discharge and drainage outlet. 

An economic alternative, semi-automatic

SAFETY

SAFETY OVER PRESSURE VALVE: PREVENTS OVER PRESSURE ABOVE THE MAXIMUM SET LIMIT.
SAFETY PRESSURE SWITCH: DISCONNECTS POWER TO THE HEATER IF THE UNIT EXCEEDS THE MAXIMUM PRESSURE LIMIT.
SAFETY THERMOSTAT AND WATER LEVEL: DISCONNECT POWER TO THE HEATER IF INSUFFICIENT WATER OR OVER TEMPERATURE.
SAFETY LID: PREVENTS THE LID BEING OPENED IF THE CHAMBER IS UNDER PRESSURE.
SAFETY INSULATION: THERMALLY INSULATED LID. 

FEATURES

External case made from stainless steel.
INCOLOY stainless steel heater elements, hard waring, resistant to corrosion and high temperatures.
RS-232 connection.
External lateral probe port.
Sterilization chamber and door made from stainless steel.
Automatic purge to atmosphere