Centrífuga refrigerada de gran capacidad y control por microprocesador para tubos, frascos y bolsas de sangre “Macrofriger-BLT-Blood”

Refrigerated high capacity microprocessor controlled centrifuge suitable for tubes, flasks and blood bags “Macrofriger-BLT-Blood”

Adjustable working temperatures from 0 to 40 ° c. (temperature at standstill -20 ° c). Capacity up to 1000 ml bottles and special for triple and quadruple blood bags of 500 to 800 ml.
Maximum capacity 12 bags.

SAFETY:

SAFETY, CONFORMS TO UNE-EN 61010-2-020.
PROTECTIVE INTERNAL CYLINDER BETWEEN THE CASE AND BOWL. THE LID CANNOT BE OPENED WHILE THE ROTOR IS SPINNING.
THE MOTOR CANNOT SPIN WHILE THE LID IS OPEN.
IMBALANCE DETECTOR, AUTOMATICALLY STOPS IF THERE IS ANY IMBALANCE. DOUBLE ELECTROMAGNETIC LID LOCK.
IF THERE IS A POWER FAILURE, A FAIL SAFE SYSTEM CAN BE USED TO OPEN THE LID.
EMERGENCY STOP BUTTON. 

COMMON FEATURES

7” TFT touch screen.
Furniture: aluminum structure in Megafriger model and in steel tube with Dur-Al sides in Macrofriger model.
Internal Stainless steel bowl.
Aluminium windbreak cover.
Safety shield: 8 mm thick steel.
Swing-out rotor, buckets and adapters made of high-density material, engraved identification unalterable.
Suspended motor with silent-block and shock absorbers that guarantee smooth, quiet operation without vibrations.
Drive force: equipped with an induction drive maintenance free and highly efficient pollution free motor.
Semi-hermetically sealed, compressor pollution free with no hydrocarbons or fluorocarbons.
Mounted on an anti vibration adjustable rubber feet.
Microprocessor digital control of all functions: speed, acceleration, brake, R.C.F., temperature, time, 10 program memory, rotor identification, operation fault alarm.
Digital indicator of the rotor and blood bag buckets operating hours. 

APPLICATIONS

Specially designed for blood bank applications, where repeatability, precision and reliability of: speed, time, temperature, acceleration and deceleration are significant factors in obtaining perfect separations of the distinct blood fractions.
Serology investigatios.
Molecular biology.
Serum separation, pharmaceutical applications. 

CONTROL PANEL

Touch screen 
1. Stopprocesskey.
2. Startprocesskey.
3. Increasevaluekey.
4. Decreasevaluekey.
5. Scheduledaccelerationramp.
6. Rotormodelinuse.
7. Reading the real time temperature. (refrigerated models only).
8. Door opening key.
9. RPM or RCF real-time reading.
10. Sequential operation.
11. Timeset-upinprocess.
12. Key to change RPM / RCF.
13. Braking ramp programmed.
14. Return key.

TECHNICAL SPECIFICATION

Maximum volume: 9600 ml
Maximum bucket volume: 1600 ml
Speed 7001490: from 300 to 4070 r.p.m.
Accuracy: ± 5 r.p.m
Resolution r.p.m.: 10 r.p.m.
Imbalance: ± 5 g
Accuracy: 4 g
Temperature range: from -20 °C to 40 °C
Temperature stability: ±4 °C
Resolution: 1 °C
Temperature range: 99 min. 59 sec.
Centrifuge radius:  290 mm
Acceleration profile: 5
Brake profile: 9 linear 6 quadratic
Voltage: 230/400 three phase
Power: 4500 VA
CódigoMax. Volume mlMax. Number of tubesHeight (exterior) cmWidth (exterior) cmDepth (exterior) cmConsumption wWeight Kg
7001490 9600 12 bags x 800 ml 105 96 96 5500 420