Banco de sangre "Blood Bank" E

Platelet bags incubated cabinets “Blood Bank” E

Exclusive design adapted to accomodate platelet bags stirrers “apl”. Forced air circulation.
Fixed temperature at +20 °c. Stability ±1,5 °c. Resolution 1 °c.
Seven days graphic temperature register.

SAFETY

LOW TEMPERATURE THERMOSTAT LIMITER.
VISUAL AND ACOUSTIC INDICATORS: EXCESSIVE LOW TEMPERATURE (+18 °C), EXCESSIVE HIGH TEMPERATURE (+24 °C), OPEN DOOR, MAINS FAILURE TEMPERATURE GRAPHIC REGISTER. 7 DAYS RECORDER, AUTONOMY 6 MONTHS.
ACOUSTIC ALARM ON ALL FAIL CONDITIONS. ACOUSTIC INDICATOR.

APPLICATIONS

Storage cabinet specially designed for platelet bags preservation. 

COMMON FEATURES

AISI 304 stainless steel external cabinet and internal lining.
Double glazed door, reversible lockable door, easy to change, with automatic feed- back mechanism and manual locking.
Hermetically sealed compressor mounted on antivibration settlement with forced type evaporator and ventilated type condenser.
Shelves and supports in steel PVC laminated.
2 power outlet base for model “Blood-Bank E” max. power 500 W
Inner air circulation for a more temperature homogeneity.
Temperature sensors located in similar conditions to platelet bags.
Circular graphic register built-in.
Microprocessor temperature control that allows configuration and visualization of the following parameters by means of the display:
- SET adjustment.
- Alarm adjustment.
- Calibration. 
CódigoCapacity litresCapacity stirrersHeight (interior) cmWidth (interior) cmDepth (interior) cmHeight (exterior) cmWidth (exterior) cmDepth (exterior) cmConsumption wPotencia compresor H.P.Nº shelvesWeight Kg
2101511 5602 x APL-54 153 54 66 213 80 79 497 3/8 4 165
2101513 1240 4xAPL-54 o 2xAPL-108 153 124 68 213 139 79 1034 5/8 8 212